Miscellaneous - EAC Meter Calibration Laboratory

 

On the 17th of January 2006 at 9:00 am electric and magnetic field measurements were performed for the EAC Meter Calibration Laboratory. The measured EAC meter calibration laboratory is located in Meteoron Street in Nicosia.

eac_meter_lab.jpg eac_meter_lab_inside.jpg


Limiting exposure values have been provided by the ICNIRP recommendations and are guidelines to many countries and organizations. The exposure limits at 50 Hz are tabulated in the following table according to the limits given by ICNIRP.

Reference levels for Occupational exposure ICNIRP

Frequency

Electric Field Strength, E (V/m)

Magnetic Field Strength, H (A/m)

Magnetic Flux Density, B (µT)

50 Hz

10000

400

500

Reference levels for General Public exposure ICNIRP

Frequency

Electric Field Strength, E (V/m)

Magnetic Field Strength, H (A/m)

Magnetic Flux Density, B (µT)

50 Hz

5000

80

100

Table 2: ICNIRP exposure limits for Occupational and General Public exposure

The obtained results for the RMS fields inside the substation are shown in table 3 and a comparison is made with the ICNIRP recommendations.


EAC Meter Calibration Laboratory: Magnetic Field Measurements

Distance from Side (m)

Maximum Magnetic Field (µT) at Z Height (m)
Percentage of ICNIRP Occupational Standard (%)

0.3m

ICNIRP %

1 m

ICNIRP %

1.75 m

ICNIRP %

Field measurements inside the substation (distance from left inside wall)

0m from Front Wall 1

0,62

0,12%

1,02

0,20%

0,76

0,15%

1m from Front Wall 1

21,43

4,29%

3,96

0,79%

0,98

0,20%

2m from Front Wall 1

2,06

0,41%

5,88

1,18%

4,09

0,82%

3m from Front Wall 1

52,13

10,43%

25,65

5,13%

4,73

0,95%

4m from Front Wall 1

50,76

10,15%

17,16

3,43%

4,98

1,00%

5m from Front Wall 1

135,20

27,04%

1788,50

357,70%

72,81

14,56%

6m from Front Wall 1

7,97

1,59%

9,99

2,00%

5,30

1,06%

7m from Front Wall 1

2,57

0,51%

3,58

0,72%

1,25

0,25%

8m from Front Wall 1

0,82

0,16%

0,81

0,16%

0,55

0,11%

9m from Front Wall 1

0,47

0,09%

0,41

0,08%

0,32

0,06%

Measurements of 1Ph Working Area

0m from Back Wall 2

7,00

1,40%

3,50

0,70%

1,22

0,24%

1m from Back Wall 2

8,01

1,60%

38,01

7,60%

9,50

1,90%

2m from Back Wall 2

7,35

1,47%

13,63

2,73%

6,22

1,24%

3m from Back Wall 2

2,03

0,41%

2,48

0,50%

2,11

0,42%

4m from Back Wall 2

0,80

0,16%

0,80

0,16%

0,85

0,17%

5m from Back Wall 2

0,89

0,18%

1,42

0,28%

1,42

0,28%

Measurements of Main Entrance

Measurement along Side 1, 30m

--

--

--

--

0,17

0,03%

Measurement along Side 2, 21m

--

--

--

--

0,11

0,02%

Measurement along Side 3, 30m

--

--

--

--

0,1

0,02%

Measurement along Side 4, 15m

--

--

--

--

0,29

0,06%

Measurements of Office 1

Point A

0,25

0,05%

--

--

--

--

Measurements of Office 2

Point A

0,19

0,04%

--

--

--

--

Point B

0,26

0,05%

--

--

--

--

Table 3: Maximum field measurements inside EAC meter calibration laboratory


Accordingly the highest Electric Field value measured within the area of the Laboratory was observed in Working Area 2 (1 Phase) with value 2.281 V/m and a percentage of 0.02% as compared to the ICNIRP Occupational standard.  This value was measured exactly in front and at the centre of the 1 phase meter calibration equipment.